Restoring factory calibration, Troubleshooting, Low temperature operation – UEi Test Instruments DMG100 User Manual

Page 2: Parts diagram

background image

DMG100 User’s Guide

9

Restoring Factory Calibration

The original factory calibration can be restored at any time by the following procedure:

1.

Turn the DMG100 power off.

2.

Hold

while pressing

. As soon as the power turns on, release

and

and then

press

quickly at least three times. The CAL indicator should illuminate, and

should show on the Main Numeric Display. If not, turn the power off again and repeat.

3.

Press

five times followed by

.

4.

The gauge will beep 5 times and return to normal operation. The DMG100 is now reset to the

original factory calibration.

Troubleshooting

Under certain conditions, the display may read

or

. Please use the table

below to determine and fix the problem:

Display

Mode

Possible Problem

Solution

Sensor Contaminated

Clean the Vacuum Sensor

Normal

Operation

Ambient Temperature too

Low

Turn the BluVac off, warm the

vacuum port with your hand,

and turn the BluVac back on.

Calibration

Sensor Contaminated

Clean the Vacuum Sensor

and Restart Calibration

Gauged Warmed too

Quickly

Restart Calibration. Allow the

Gauge to Warm Slowly

Calibration

Gauge Disturbed During

Calibration

Restart Calibration. Leave

the Gauge Undisturbed

During Warming Phase.

Low Temperature Operation

The DMG100 can operate accurately at temperatures as low as 10°F (-12°C). While operating below

freezing (32°F/0°C), the display update rate will slow from 3.5 readings every second to one reading

every two seconds.

A standard alkaline battey may not provide acceptable battery life at temperatures below 0°C (32°F).

For low temperature operation, a 9V Lithium battery is recommended.

DMG100 User’s Guide

2

Parts Diagram