8 specifications, Specifications, E-m-hc2-aw-usb-v1_11 – ROTRONIC HC2-AW-USB User Manual
Page 17: 8specifications

E-M-HC2-AW-USB-V1_11
Rotronic AG
Bassersdorf, Switzerland
Document code
Unit
Water Activity and Temperature Probe
HC2-AW-USB: User Guide
Instruction Manual
Document Type
Page 17 of 18
Document title
© 2011-2012; Rotronic AG E-M-HC2-AW-USB-V1_11
8
Specifications
General
Device type
Humidity temperature probe
Mechanical configuration
See models
Power supply and connections
Supply voltage (VDD)
5 V (supplied from a USB port to which the probe is connected)
Nominal current consumption
< 25 mA
Humidity measurement
Sensor
ROTRONIC Hygromer
®
WA-1
Measuring range
0…100 %RH (0.000 to 1.000 aw)
Measurement accuracy at 23 °C
±0.8 %RH (w. standard adjustment profile)
Repeatability
0.3 %RH
Long term stability
< 1 %RH / year
Temperature measurement
Sensor
Pt100 RTD, IEC 751 1/3 class B
Measuring range
-40…85 °C (-40…185 °F)
Measurement accuracy at 23 °C
±0.1 °C
Repeatability
0.05°C
Long term stability
< 0.1°C / year
Start-up time and data refresh rate
Start-up time
1.5s (typical)
Data refresh rate
1.0s (typical)
Digital interface
Interface type
USB 2.0
Signal cable length
3 m (9.8.4 ft)
General specifications
Housing material
Barrel: aluminum (top) , stainless steel (bottom)
Dust filter material
Stainless steel wire mesh
Physical dimensions
60 mm (H) x 68 mm (diameter)
Probe cable length
0.95 m (37.5”)
Weight
approx. 541 g (1 lb 3.1 oz)
Conformity to standards
CE / EMC immunity
EMC Directive 2004/108/EG: EN 61000-6-1: 2001, EN 61000-6-2: 2005
EN 61000-6-3: 2005, EN 61000-6-4: 2001 + A11
No measurement is possible in the presence of electromagnetic emissions
Solder type
Lead free (RoHS directive)
FDA / GAMP directives
compatible
Environmental limits
Storage and transit
-40…+85 °C / 0…100 %RH, non condensing