Specifications
Internal Temperature Channels
Range: -20 to +60°C
Resolution: 0.1°C
Accuracy: ±0.5°C (0 to +50°C)
Thermocouple Channels
Types: J, K, T, E, R, S, B, and N
Connection: Female sub-minature (SMP)
Cold Junction Compensation: Automatic, based on internal channel
Maximum Thermocouple Resistance: 1000
Ω
< 100
Ω
recommended
Thermocouples:
Type
Range
Resolution
Accuracy
J
-200 to +760°C
0.1°C
± 0.5°C
K
-200 to +1370°C
0.1°C
± 0.5°C
T
-200 to +400°C
0.1°C
± 0.5°C
E
-200 to +980°C
0.1°C
± 0.5°C
R
-50 to +1760°C
0.1°C
± 2.0°C
S
-50 to +1760°C
0.1°C
± 2.0°C
B
+50 to +1820°C
0.1°C
± 2.0°C
N
-200 to +1300°C
0.1°C
± 0.5°C
Dot-Matrix LCD
Dimensions: 2.5” x 1.375” (63mm x 35mm)
Text: Configurable channel text size
Indicators: Power, status, memory
Backlight: Configurable with auto shut-off and contrast adjustment
Start/ Stop Modes
Software programmable start time and date, up to six months in
advance, or manual (push-button) start. Programmable stop time.
Memory
500,000 readings per channel with all channels enabled;
2,000,000 total readings; software configurable memory wrap
Reading Rate