www.kamansensors.com

PART NO: 860527-001

Last Revised 8/23/2013

- 5 -

While pushing down on the .375” Calibration Plate, tighten the wing nuts until the Calibration plate is
firmly seated on the Base Plate. Check to make sure both of the legs on the Calibration Plate are in full
contact with the Base Plate.

Momentarily depress the calibration push button. The unit will tare to .375” NOTE: THE TARE
FUNCTION CAN ONLY BE USED AT THE .375” GAP.

General Care and Maintenance


If the unit becomes soiled use a normal spray type household cleaner and wipe it down with a clean cloth.

A light lubricating oil should be applied to the stainless body where the target springs ride.

A light lubricating oil can be applied to the outside of the target springs if desired.

System

Specifications

For

Assistance

Kaman Precision Products

Input

power:

110VAC

3730

Sinton

Road

#100

Calibrated range: 0.250” – 0.750”

Colorado Springs, CO 90807

Resolution

0.001”

1-800-562-6267

1-719-635-6957
[email protected]
www.kamansensors.com