Analyzer specification – Dwyer 1207A User Manual
Page 28

1207A manual
Page 28
12. ANALYZER SPECIFICATION
(NOTE MAY BE SUBJECT TO CHANGE)
Parameter
Resolution
Accuracy
Range
Temp Measurement
Flue Temperature
Inlet Temperature
(Internal Sensor)
Inlet Temperature
(External Sensor)
0.1
°C/F
0.1
°C/F
0.1
°C/F
+2.0
°C
+0.3% reading
+1.0°C
±0.3% reading
+2.0
°C
+0.3% reading
0-600
°C / 32-1112
o
F
0-50°C / 32-122
o
F
0-600
°C / 32-1112
o
F
Pressure Measurement
0.1 Pa
0.1 Pa
1 Pa
0.01 hPa
+0.5 Pa
+3 Pa
+3% of reading
+3% of reading
+20 Pa
+100 Pa
+2000 Pa
+80 hPa
Gas Measurement
Oxygen
Carbon Monoxide (CO)
(low range)
Carbon Monoxide (CO)
(high range)
Nitric Oxide (NO)
(low range)
Nitric Oxide (NO)
(high range)
Carbon Dioxide
*2
Efficiency
*2
Excess Air
*2
0.1%
1 ppm
1 ppm
1 ppm
1 ppm
0.1%
0.1%
0.1%
+0.2%
*1
+10 ppm <100 ppm
*1
+5% reading
+10 ppm <100 ppm
+5% reading >100 ppm
+2 ppm <30 ppm
*1
+5 ppm > 30 ppm
+5 ppm <100 ppm
+5% reading >100 ppm
+0.3% reading
+1.0% reading
+0.2%
0-21%
0-2000 ppm nom
4000 ppm max
for 15 minutes
0-4,000 ppm
0 to 100 ppm
0 to 1000 ppm
0-30%
0-99.9%
0-250%