Analyzer problem solving – Dwyer 1207A User Manual
Page 26

1207A manual
Page 26
10. ANALYZER PROBLEM SOLVING
If any problems are not solved with these solutions, contact us or an authorized repair center.
Fault symptom
Causes / Solutions
• Oxygen too high
• CO2 too low
• Air leaking into probe, tubing, water trap, connectors
or internal to analyzer.
• Oxygen cell needs replacing.
• Oxygen reading (- - - -)
• CO reading (- - - -)
• Analyzer was stored in a cold environment and is not
at normal working temperature.
• Oxygen cell or CO sensor needs replacing.
• Pump is switched off.
• Batteries not holding charge
• Analyzer not running on mains
adapter
• Batteries exhausted.
• AC charger not giving correct output.
• Fuse blown in charger plug.
• Analyzer does not respond to flue
gas
• Particle filter blocked.
• Probe or tubing blocked.
• Pump not working or damaged with contaminants.
• Net temperature or Efficiency
calculation incorrect
• Ambient temperature set wrong during Automatic
Calibration.
• Flue temperature readings erratic • Temperature plug reversed in socket.
• Faulty connection or break in cable or plug.
• T flue or T nett displays (- - - -)
• Probe not connected.
• X-Air, EFF, COa or CO2 display
(- - - -)
• Oxygen reading is above 18%.
• Analyzer just continually beeps
• Turn dial back to MENU and press
.
• BAT only shows 65 with fully
charged NiMh batteries fitted
• This is not a problem and is to be expected as NiMh
batteries only deliver 1.25 V per cell whereas
Alkalines deliver 1.5 V per cell. Fresh alkalines might
give a BAT value of 80 or so.