beautypg.com

Calibration, Calibrating the 1017, Zero adjustment – Time Electronics 1017 DC Multifunction V_I_R Calibrator User Manual

Page 8: C alibration

background image

8

1017 Technical Manual

8

4. Calibration

4.1. Calibrating the 1017

The 1017 can be calibrated with a high accuracy D.M.M. with D.C. voltage, current and
resistance ranges with a specification of at least 4 times greater than that of the 1017’s
specifications.

Calibration is best carried out on fully charged batteries without the mains supply connected
to ensure that no mains supply interference takes any effect on the readings.


4.2. Zero Adjustment

Range

Allowance

Adjust Pot


1V

6µV

1V ZERO ON MODULE