Calibration, Calibrating the 1017, Zero adjustment – Time Electronics 1017 DC Multifunction V_I_R Calibrator User Manual
Page 8: C alibration
8
1017 Technical Manual
8
4. Calibration
4.1. Calibrating the 1017
The 1017 can be calibrated with a high accuracy D.M.M. with D.C. voltage, current and
resistance ranges with a specification of at least 4 times greater than that of the 1017’s
specifications.
Calibration is best carried out on fully charged batteries without the mains supply connected
to ensure that no mains supply interference takes any effect on the readings.
4.2. Zero Adjustment
Range
Allowance
Adjust Pot
1V
6µV
1V ZERO ON MODULE