Specifications – Time Electronics 1017 DC Multifunction V_I_R Calibrator User Manual
Page 4
4
1017 Technical Manual
4
1.2. Specifications
Outputs:
0 – 100V in 5 Ranges
0 – 100mA in 1 Range
0 – 10k
Ω in 1 Range
Voltage
Range:
0 – 9.99999mV in 10nV steps
0 – 99.9999mV in 100nV steps
0 – 999.999mV in 1uV steps
0 – 9.99999V in 10uV steps
0 – 99.9999V in 100uV steps
Accuracy:
10mV range
0.02% of setting +/- 0.005% of range
100mV range
0.01% of setting +/- 0.004% of range
1V range
0.005% of setting +/- 0.002% of range
10V range
0.005% of setting +/- 0.002% of range
100V range
0.01% of setting +/- 0.004% of range
Output Resistance:
100mV to 10V
<200m
Ω
10mV and 100V
<1
Ω
Output Current:
100mV to 10V
150mA
100V
10mA
10mV
Limited by 1
Ω Output resistance
All specification +/- 3uV.
Current
Range:
0-99.9999mA
in 100nA steps
Accuracy:
100mA range
0.02% of setting and 0.004% of range
Drive Voltage:
10V Maximum
Resistance
Range:
0 – 9.99999k
Ω in 10mΩ steps
Accuracy:
0.05% of full scale
Power Rating:
250mW per Resistor
End Resistance:
< 200m
Ω
Deviation Control (Only on Voltage and Current)
Range:
0.001% to 0.999% in 0.001% steps
Accuracy:
0.5% Accuracy
Stability
Less than 10ppm/°C on all ranges.
Less than 5ppm/Day at constant temperature
Less than 15ppm/90 days
Less than 25ppm/Year
Output Noise
Output Noise (0.1Hz to 10Hz) - Typically 3ppm.
Power Supply
The 1017 can be used continuously from a 220V / 240V 50 / 60Hz (110V / 120V to order)
mains supply or from the internal rechargeable Ni-Cad battery pack. A front panel meter
indicates the state of charge at all times.