Thermo Fisher Scientific ECCS10-0-1S Conductivity/Resistivity Sensors User Manual
Page 3

www.thermofisher.com
Demineralised water
1 – 80
μS/cm
0.01 – 1 M
Ω-cm
Mountain water
10
μS/cm 0.1
M
Ω-cm
Drinking water
0.5 – 1 mS/cm
1- 2 k
Ω-cm
Wastewater
0.9 – 9 mS/cm
0.1 – 1 k
Ω-cm
KCl solution (0.01 M)
1.4 mS/cm
0.7 k
Ω-cm
Potable water maximum
1.5 mS/cm
0.1 k
Ω-cm
Brackish water
1 – 80 mS/cm
0.01 – 1 k
Ω-cm
Industrial process water
7 – 140 mS/cm
rarely stated
Ocean water
53 mS/cm
rarely stated
10% NaOH
355 mS/cm
rarely stated
31% HNO
3
865 mS/cm
rarely stated
WIRING CONNECTIONS FOR EUTECH INSTRUMENTS CONDUCTIVITY/ RESISTIVITY CELLS
RED
BROWN
GREEN
WHITE
BLACK
SHIELD
To Electrode
Conductivity / Resistivity
Cells
Eutech
Alpha
CON1000
Eutech
Alpha
CON2000P
Alpha
CON2000W
Eutech
Alpha
CON/TDS200
Eutech
AlphaRES100
0
Eutech
AlphaCON500
(for CS10-0-1S
& CS10-1-0S
only)
Eutech
AlphaCON550
AlphaCON560
(for CS10-0-1S
& CS10-1-0S
only)
Model
Color Code
Pin No.
Pin No.
Pin No.
Pin No.
Pin No.
Pin No.
RED 19 25 13 19
5
5
GREEN 20
24
15
20
7
7
BROWN 18
26
14
18
6
6
BLACK 21 15-16 16
21
4
3
WHITE 22 17-18 17
22
3
2
ECCS10-0-01S
ECCS10-0-01T
ECCS10-0-1S
ECCS10-1-0S
**SHIELD
Gnd Tab
22
20 or 21
Gnd Tab
7
7
NOTE **: It is important to connect the SHIELD wire to the pin no as stated above to eliminate possible oscillation of
the measured Conductivity/TDS/Resistivity and/or temperature reading due to electrical noise or interference.