Demineralised water

1 – 80

μS/cm

0.01 – 1 M

Ω-cm

Mountain water

10

μS/cm 0.1

M

Ω-cm

Drinking water

0.5 – 1 mS/cm

1- 2 k

Ω-cm

Wastewater

0.9 – 9 mS/cm

0.1 – 1 k

Ω-cm

KCl solution (0.01 M)

1.4 mS/cm

0.7 k

Ω-cm

Potable water maximum

1.5 mS/cm

0.1 k

Ω-cm

Brackish water

1 – 80 mS/cm

0.01 – 1 k

Ω-cm

Industrial process water

7 – 140 mS/cm

rarely stated

Ocean water

53 mS/cm

rarely stated

10% NaOH

355 mS/cm

rarely stated

31% HNO

3

865 mS/cm

rarely stated

WIRING CONNECTIONS FOR EUTECH INSTRUMENTS CONDUCTIVITY/ RESISTIVITY CELLS

RED

BROWN

GREEN

WHITE

BLACK

SHIELD

To Electrode

Conductivity / Resistivity

Cells

Eutech

Alpha

CON1000

Eutech

Alpha

CON2000P

Alpha

CON2000W

Eutech

Alpha

CON/TDS200

Eutech

AlphaRES100

0

Eutech

AlphaCON500
(for CS10-0-1S

& CS10-1-0S

only)

Eutech

AlphaCON550
AlphaCON560
(for CS10-0-1S

& CS10-1-0S

only)

Model

Color Code

Pin No.

Pin No.

Pin No.

Pin No.

Pin No.

Pin No.

RED 19 25 13 19

5

5

GREEN 20

24

15

20

7

7

BROWN 18

26

14

18

6

6

BLACK 21 15-16 16

21

4

3

WHITE 22 17-18 17

22

3

2

ECCS10-0-01S
ECCS10-0-01T

ECCS10-0-1S
ECCS10-1-0S

**SHIELD

Gnd Tab

22

20 or 21

Gnd Tab

7

7


NOTE **:
It is important to connect the SHIELD wire to the pin no as stated above to eliminate possible oscillation of