TROUBLESHOOTING HINTS
Symptom Possible Causes
Next Step
Out of Range defective meter
check meter with shorting
Reading
strap (see meter
instruction manual)
defective electrode
check electrode operation
electrodes not
plug electrode into position
plugged
in
properly
and
reset
meter
reference electrode
clean junction
junction clogged
static
electricity wipe
plastic
meter
face
with
detergent solution
meter
or
stirrer ground
meter
or
stirrer
not
grounded
Noisy or Unstable defective meter
check meter with
Readings (readings
shorting strap
continuously or
rapidly changing) air bubble on membrane
remove bubble by
re-dipping electrode
ISA not used
use recommended ISA
meter or stirrer
ground meter or
not grounded
stirrer
defective electrode
replace electrode
electrode exposed
soak electrode in
to interferences
chloride standard
Drift (reading
samples and standards
allow solutions to come to
slowly changing at different temperatures
room temperature before
in one direction)
measurement
incorrect reference
use recommended
filling solution
filling solution in outer
of reference electrode
membrane
dirty
or
etched
polish
membrane