9.7 3.3 AtC (Auto Temp Compensation)—Con & TDS only

Press

.

Press

or

to select “Yes” or “No”.

Press

to confirm.

9.8 3.4 tdS (TDS factor)—Con & TDS only

Press

.

Press

or

to select the desired TDS factor (.40 to

1.00).

Press

to confirm.

9.9 3.5 t.CO (Temperature Coefficient)—Con & TDS only
The temperature coefficient is the amount of change in conductivity per

degree temperature (% per ºC). The PC 700 is factory set to a

temperature coefficient of 2.1 % per ºC. For most applications this will

provide good results. The meter allows adjustment from 0.0 to 10.0.
TIP: Select 0.0% for uncompensated measurements. The temperature

will be measured by the electrode and displayed in measurement

mode—without compensation.

See Section 11

– Calculating Temperature Coefficients.

Press

.

Press

or

to select the desired value (0.00 to 10.0).

Press

to confirm.