13. MIN AND MAX FUNCTION

Press the min/max key to toggle between

the minimum and maximum readings. The

minimum and maximum reading function

is ideal for monitoring unattended

operations while continually displaying

every temperature change that occurs.

The minimum and maximum values are

sensed and automatically stored.

To exit and clear this function, press the
min/max

key for 3 seconds.

14. AUTO OFF FUNCTION


The thermometer has a default auto off
function of 17.5 minutes.

To deactivate this function, press and hold
the min/max key while turning on the
meter using the on/off key.
“A.OFF nO” will appear indicating that the
auto off function is disabled.

15. CALIBRATION

The thermometer is factory calibrated and

does not require calibration before use.

The Calibration function allows single point

calibration of the thermometer, at 0°C

(32°F) to compensate for thermocouple

off-set error. It is NOT necessary to

perform a field calibration to obtain

specified meter accuracy. Use the field

calibration feature to improve

thermometer/probe accuracy or to