Thermo Fisher Scientific Thermo Scientific Temp 16 User Manual
Page 8

4
3. SPECIFICATIONS
Platinum RTD Probes
100 ohm platinum: alpha = 0.003850
–200 °C to 850 °C (–392 °F to 1562 °F)
Out of range display: “Ur”, “Or”, “OPEN”
Range Accuracy
–200.0 to –100.0 °C
(–392 to –148 °F)
±2.0 °C
(±4.0 °F)
–99.9 to 199.9 °C
(–148 to 392 °F)
±0.2 °C
(±0.4 °F)
200.0 to 850.0 °C
(392 to 1562 °F)
±2.0 °C
(±4.0 °F)
Resolution
0.1° (–200.0 to 850.0 °C);
0.1° (–392.0 to 999.9 °F);
1° (1000 to 1562 °F)
Display
4-digit LCD (¼ X ½" digits);
2
1
/
3
X 1 ½" backlit viewing
area
Display Update
Rate
0.5 sec per update.
Input
One probe with 3 pin
circular connector
(Switchcraft TA3F)
Input Protection
10V rms
Battery
Three AA, 1.5V alkaline
Battery Life
700 hours continuous
Auto Shutoff
17 minutes after last
keypress.
Stated Accuracy
18 to 28°C
(64 to 82°F)
Useful Range
0 to 40°C
(32 to 104°F)
Storage
-40 to 65°C
(-40 to 149°F)
Humidity
10% to 90%
(non-condensing)
Dimensions
Without armour
With armour
(L x W x H )
175 mm x 97 mm x 42 mm
180 mm x 102 mm x 52 mm
- PCTestr 35 (2 pages)
- pHScan BNC (3 pages)
- pHScan 3/3+ (5 pages)
- pHTestr 1 (3 pages)
- pHTestr 10/20/30/10 BNC/Spear (2 pages)
- ORPTestr 10/10 BNC (2 pages)
- EC/TDS/SaltTestr 11 (4 pages)
- EC/TDS/SaltTestr (2 pages)
- ECScan High/Low & TDScan High/Low (9 pages)
- SaltTestr (2 pages)
- EcoTestr pH 2 (2 pages)
- EcoTestr pH 1 (2 pages)
- EcoTestr EC High (2 pages)
- EcoTestr EC Low (2 pages)
- EcoTestr TDS High (2 pages)
- EcoTestr TDS Low (2 pages)
- EcoTestr Salt (2 pages)
- Eutech pH 5/6 Plus & Ion 6 Plus (New version R1.1, SN >797406) (23 pages)
- Eutech pH 5/6 Plus & Ion 6 Plus (Old version EP6, SN <797406, discontinued) (23 pages)
- Eutech COND/TDS/Salt 6 Plus (40 pages)
- Eutech DO 6 Plus (48 pages)
- EcoScan pH/Ion 5 & 6 (27 pages)
- EcoScan CON 6 & TDS 6 (56 pages)
- EcoScan CON 5 & TDS 5 (18 pages)
- EcoScan Salt 6 (40 pages)
- EcoScan DO 6 (80 pages)
- CyberScan pH 10/pH 100 (67 pages)
- CyberScan pH 11/pH 110 (76 pages)
- CyberScan CON 10/CON 100/CON 200 (62 pages)
- CyberScan CON 11/CON 110 (80 pages)
- CyberScan DO 110 (60 pages)
- CyberScan PCD 650 (127 pages)
- CyberScan CON 400/410 (For units manufactured before March 2010, discontinued) (60 pages)
- CyberScan CON 400 (For units manufactured from March 2010 onwards) (60 pages)
- CyberScan pH 300/310 (52 pages)
- CyberScan DO 300 (60 pages)
- CyberScan PC 300 (72 pages)
- CyberScan PD 300 (76 pages)
- CyberScan PC 10 (31 pages)
- C401 Colorimeter (64 pages)
- TN100 Turbidimeter (31 pages)
- RS232C Interface Adapter (9 pages)
- Thermo Scientific Temp 360 (44 pages)
- Thermo Scientific Temp 340 (40 pages)
- Thermo Scientific Temp 300 (32 pages)