Electrode replacement, Applications – Thermo Fisher Scientific ECScan High/Low & TDScan High/Low User Manual
Page 9

A
B
REMOVE THE OLD ELECTRODE
REPLACE THE ELECTRODE
(Picture B)
(Picture A)
Electrode Replacement
You can replace the electrode module at
the fraction of the cost of a new tester.
When the tester fails to calibrate or gives
fluctuating readings in calibration
standards, you need to change the
electrode.
1.
With dry hands, grip the ribbed tester
collar with electrode facing you. Twist the
collar counter clockwise (see picture A).
Save the ribbed tester collar and O-ring
for later use.
2.
Pull the old electrode module away from
the tester.
3.
Align the four tabs on the new module
so that they match the four slots on the
tester (see picture B).
4.
Gently push the module onto the slots
to sit it in position. Push the smaller O-
ring fully onto the new electrode module.
Push the collar over the module and thread
it into place by firmly twisting clockwise.
Water quality testing • water and
Wastewater treatment • hydroponics
• labs • boilers • factories
• ecology studies • and more!
Applications
Tester
TDScan TDScan
ECScan ECScan
Low
High
Low
High
Range
0 to
0 to
0 to
0 to
1990
10.00
1990
19.90
ppm
ppt
µS/cm
mS/cm
Resolution
10
0.10
10
0.10
ppm
ppt
µS/cm
mS/cm
Accuracy
±1% Full Scale
TDS factor 0.4 to 1.0 selectable
-
Calibration
200 to
2 .00 to
200 to 2.00 to
Standard
1990
10.00
1990
19.90
Range
ppm
ppt
µS/cm
mS/cm
Calibration
1 point (calibration range is ±30%
of factory default parameter)
ATC
0 to 50 °C
Temp
2% per °C
Coefficient
Operating
0 to 50 °C
Temp
Power
Four X 1.5V
Battery
(Type: A76, LR 44 or V136A);
> 140 hrs continuous use
Dimension;
Tester: 16.5 cm x 3.8 cm; 90 g
Weight
Boxed: 22 cm x6 cm x 5 cm; 170 g
•
Push-button calibration
•
Fast, stable, repeatable readings
•
Available through authorized distributors
S P E C I F I C A T I O N S
Lanyard
connection
Replaceable
Sensor
Twist-off
Battery
Compartment
Manufactured By:
Eutech Instruments Pte Ltd
Blk 55, Ayer Rajah Crescent,
#04-14/24, Singapore 139949
Tel: (65)778 6876 Fax: (65)773 0836
E-Mail: [email protected]
Web-Site: http://www.eutechinst.com
Made in Singapore
68X247828 11/00 Rev 0