Waterproof pHTestr 1

Microprocessor-based Pocket pH Tester

68X247716 2/05 Rev 4

MAINTENANCE IS EASY:

Rinse the

electrode in tap water after use. In harsh
samples, take readings quickly, rinse electrode
immediately afterwards with tap water to
remove residues and prevent electrode
contamination. To maximize electrode life, place
a small piece of clean cloth or sponge in the
cap, moisten with tap water (NOT DE-IONIZED
WATER), and replace cap.

CHANGING BATTERIES:

Twist open the

battery compar tment lid. Remove old
batteries and replace with fresh ones noting
polarity as shown in battery compartment.
Re-calibrate after battery change.

USEFUL NOTE:

• To avoid cross contami-

nation, rinse between samples and buffer
with tap water • Calibration should be done
regularly or daily to ensure good tester
accuracy.

ERROR MESSAGES:

E1 - Weak batteries.

Need replacement. E2 - Wrong or bad buffer
value (out of range) or the electrode is failing.
OR - Over range signal or electrode is not in
contact with sample solution.

BEFORE USE:

DO NOT BE ALARMED if

white crystals form around the cap. This

is normal.

CONDITIONING:

Condition the unit before

first use. Remove cap, immerse electrode

in tap water for 1 hour. This activates

electrode and dissolves crystals.

CALIBRATION:

Select pH 7 buffer for

general testing or pH 4 or 10 buffer

if you are measuring acidic/basic

samples. Press ON/OFF button to power

on. Immerse electrode in chosen

buffer about 2 cm deep and stir gently.

Wait for displayed value to stabilize

at or near the pH buffer chosen.

Next, press the CAL button to enter

calibration sequence. When dis-

play flashes continuously, press

HOLD/CON button to confirm.

You have successfully calibrated

the instrument.

pH TESTING:

• Remove cap

and press ON/OFF button on
the keypad to turn on the
pHTestr 1 • Dip the electrode

about 2 cm into the test solution

• Stir once, let the display

stabilize. Note the pH • Press

HOLD/CON button if you wish

to hold the reading. Press again

to release • Press the ON/OFF

button to shut off • Tester au-

tomatically shuts-off after 8.5

minutes to conserve batteries!

Keypad