Waterproof pHScan BNC
Microprocessor-based Pocket pH Tester

Connecting your electrode:

To connect the tester to your electrode, align the

posts of the tester's BNC connector with the slots

on the electrode connector. Push together and twist

the electrode connector 1/2 turn until it clicks into
place.

Calibration:

Calibration should be done regularly, typically

every day that the tester is used. You can calibrate

up to three point. Choose buffer sets as described

above right.

1. Press ON/OFF button to switch unit on.

2. Dip electrode about 2 cm into buffer.

3. Press CAL button to enter Calibrate (CAL)

mode. 'CAL' flashes on the display. Then a

pH value close to pH buffer value will flash

repeatedly.

4. After at least 3 minutes, press

the HOLD/CON button to confirm

calibration. The display will show

'CO' and then switch to the buffer

value reading.

5. Repeat with other buffers if

necessary. Rinse electrode in tap

water before dipping into next buffer.

Calibration Troubleshooting:

Failure to press HOLD/CON to
confirm calibration (step 4 above).
Pressing the CAL button will resume

measuring mode but will not enter
the calibration value.

Insufficient sampling time. If the

tester does not have a long enough

exposure to the buffer, a stable

calibration point will not be reached.

Wait at least 3 minutes before

pressing HOLD/CON.

Failure to re-hydrate the electrode. A

dry electrode will give fluctuating

reading while it re-hydrates in a buffer,

causing errors.

Buffer Set selections:

To choose USA (pH 4.01, 7.00, 10.01) or NIST
(pH 4.01, 6.86, 9.18) buffers:
1. With tester off, hold HOLD/CON key while

pressing ON/OFF key.

2. Press HOLD/CON key again.
3. Press CAL key to toggle between USA and NIST

buffer sets.

4. Press HOLD/CON to confirm.

pH Testing:

1. Remove cap or storage bottle from the

electrode and press the ON/OFF button to
switch tester on.

2. Dip the electrode about 2 cm into the test

solution. Stir once and let the reading stabilize.

3. Note the pH or press HOLD/CON button to

freeze the reading. Press HOLD/CON again to
release the reading.

4. Press ON/OFF to turn off tester. If you do not

press a button for 8.5 minutes the tester will
automatically shut off to conserve batteries.

Instrument Maintenance:

• Rinse the electrode with tap water or electrode

storage solution after each measurement.

• Periodic soaks in warm pH 4 buffer will help

remove contaminants.

• When finished using electrode, fill electrode cap

with tap water or electrode storage solution and
replace cap.