EM101 LOW FLOW
MAGNETIC FLOW METER
INSTRUCTIONS
www.seametrics.com
1 9 0 2 6 7 2 N D AV E S O U T H , K E N T, WA 9 8 0 3 2 U S A
( P ) 2 5 3 . 8 7 2 . 0 2 8 4 ( F ) 2 5 3 . 8 7 2 . 0 2 8 5
W W W. S E A M E T R I C S . C O M 1 . 8 0 0 . 9 7 5 . 8 1 5 3
LT - 1 1 3 1 4 - A
EM101 TROUBLESHOOTING GUIDE
Problem Possible Causes Possible Solutions
No Display
Flow rate always reads “0”
Flow rate incorrect
No power
Blown Fuse
Loose ribbon cable
Reversed flow direction
Missing ground wire
Empty pipe
Flow rate below minimum
Loose wiring or incorret wiring
Fluid conductivity <20 uS/cm
Missing or incorrect ground wire
Fluid conductivity <20 uS/cm
Empty pipe
Check for 110vac
Test fuse, replace if bad
Check robbon cable commections
Reverse flow connections
Install ground wire
Install meter in vertical position
Usr next smaller flow meter
Check connections on display board
Select other flow meter
Check for proper grounding
Select other flow meter
Install meter in vertical position