beautypg.com

Pyromation 28-02700-T User Manual

Page 14

background image

12

FIELD CALIBRATION LOCKOUT
AND RE-ENABLE

The calibration lockout feature, prevents any field
calibration changes. The lockout remains in effect
until a lockout re-enable has been performed. Use
the following procedures to lockout or re-enable the
field calibration operation.

Lockout Procedure

1. Turn the thermometer off.

2. Simultaneously press and hold the CAL and the

°C/°F keys down and momentarily press the ON/
OFF
key. Continue to hold the CAL and °C/°F
keys for at least 5 seconds.

Re-Enable Procedure

1. Turn the thermometer off.

2. Simultaneously press and hold the HOLD and

the CAL keys down and momentarily press the
ON/OFF key. Continue to hold the HOLD and
CAL keys until the display blanks.

This manual is related to the following products: