Metex 2200R ORP Analyzer/Controller User Manual
Page 25

Model 2200R
Page 23
8.0
DIAGNOSTICS
8.1 Description
8.1.1
The 2200R has diagnostic features which alerts the operator to invalid entries and
memory loss. Invalid entries are indicated by the flashing of the appropriate menu
LED. The flashing will commence after RUN is pressed and will continue until the
errors are corrected. Memory loss is indicated by the flashing of TEST and by the
alarm relay if enabled by DIP Switch No. 5 of Bank S1. Refer to Section 4.7.3.
8.2
Invalid
Calibration
3.5.3
Invalid calibration will be indicated in the following instances:
• When the difference between the two calibration buffers is less than 10% of the
measuring scale span. To correct, enter the menu and calibrate with more
appropriate buffers. Refer to Section 5.2.
• When the input is out of range. This could occur if you had attempted to calibrate
with a buffer which was outside the range of the instrument.
8.3 Invalid Output
8.3.1
Invalid output will be indicated if the expanded range is less than 10% of full scale.
To
correct,
refer
to
Section
5.4.
8.4 Invalid Alarm Points
8.4.1
Invalid Alarm will be indicated if the Low Alarm is set higher than the High Alarm.
To correct, refer to Section 5.10 and 5.11.
8.5 Invalid Setpoint
8.5.1
The deadband setting must be lower than the Setpoint if you are controlling rising
ORP, and higher than the Setpoint if you are controlling falling ORP. If you have
not made these settings correctly, invalid entry will be indicated by a flashing LED.
To correct, refer to Section 5.5 and 5.8.