Martel Electronics DPM970 User Manual

Dpm 970, Dimensions, Digital ac voltmeter module

background image

DIMENSIONS

All dimensions in mm (inches)

REAR
VIEW

ACV ACI COM COM V+

V-

LP-

LP+

V+

DP1

DP2

DP3

TST

HLD

ALM

XDP

RB

Required F.S.R.

RA2

RB

RA1

200mV

0R

0R

Open

2V

910k

100k

20V

1M

10k

200V

1M

1k

2kV (factory default)

464K

536K

100R

2mA (factory default)

-

-

100R

20mA

-

-

10R

200mA

-

-

1R

2A

-

-

0.1R

20A

-

-

0.01R




➀ ➁




➂ ➃

0R
0R
0R

DPM 970

Digital AC Voltmeter Module

19mm (0.75“) Digit Height

Screw Terminal Connections

Digital Hold

500V a.c. (2kV ) Full Scale Reading

Measures Voltage and Current

LED Backlighting

A 3 ½ digit LCD panel meter, featuring 19mm (0.75“) digit height and LED backlighting, the DPM 970 is a true RMS voltage and current meter with 500V a.c.
(2kV ) and 2mA full scale readings as standard. By changing the value of scaling resistors or adding a shunt, this module will measure most a.c. voltages and
currents. The DPM 970 features auto-zero and suppression of d.c. offset voltage. The digital hold function freezes the last displayed reading. Calibration is
achieved via two multi-turn potentiometers for offset and scale. Connection to the meter is via screw terminals. The module is housed in a robust carrier
which can be bolted in place or panel mounted using the bezel, window and clips provided.

* To ensure maximum accuracy, re-calibrate periodically.
** Maximum safe input voltage across the ACV and COM terminals is 500V

.

RMS

RMS

If voltages to be measured exceed 500V

, then fit scaling resistors

externally to the module.

Specification

Min.

Typ.

Max.

Unit

Accuracy (overall error) *

0.5

1.5

% (±2 count)

Sample rate

3

samples/sec

Operating temperature range

0

50

°C

Temperature s

200

ppm/°C

Frequency range (input voltage)

15

15k

Hz

Input impedance (2kV range)

1

M

DC Supply voltage

7.5

9

14

V d.c.

Supply current (Backlighting off)

1.5

mA d.c.

Supply current (Backlighting on)

50

mA d.c.

Full scale reading **

0

1999

V

Resolution

1

V

RMS

RMS

tability

@ 5Vd.c.

Stock Number

Standard Meter

DPM 970

61.5 (2.42)

51.0 (2.01)

Ш5.7

(0.22)

24.5

(0.96)

24.0

(0.94)

39.0

(1.54)

Ш3.0

(0.12)

71.5 (2.81)

d

76.0 (2.99)

72.5 (2.85)

40.5

(1.59)

44.0

(1.73)

24.0

(0.94)

51.0 (2.01)

a

b c

a.

1.0 (0.04)

b.

2.0 (0.08)

c.

6.5 (0.26)

d. 11.5 (0.45)

51 (2.01)

24

(0.94)

Display in Test Mode

WEBSITE: http://www.lascarelectronics.com/

Panel cut-out 72 x 40 (2.83 x 1.57)
Maximum panel thickness 3mm (0.12).

SCALING

The value of the three scaling resistors RA1, RA2 and RB may be
altered in order to change the full scale reading (F.S.R.) of the meter
- see table.

The meter will need re-calibration by adjusting the

CAL potentiometer.

To ensure that the meter reads "000" with zero input

voltage/current, short circuit ACV and COM and adjust the ZERO
potentiometer.

NOTE


Use inputs ACV and COM for measuring voltage.
For measuring voltages above 500V RMS, external scaling
resistors must be used.
Use inputs ACI and COM for measuring current.
We recommend the use of a current transformer when
measuring in excess of 10 Amps. Consult current transformer
manufacturer’s data for connections and recommendations.

Other scaling factors can be achieved via the correct selection of
RA1, RA2 and RB values.

CAL ZERO

RA1

RA2