Wire functions, Safety, Various operating modes – Martel Electronics QM130M User Manual
Page 2: Power supply, Scaling

WIRE FUNCTIONS
RED
V+
Positive power supply connection.
BLACK
0V
Negative power supply connection.
WHITE
Vin
Positive measuring input with reference to 0V.
Range from 0 to 1V d.c.
SAFETY
To comply with the Low Voltage Directive (LVD 93/68/EEC), input voltages to the module’s pins must not exceed 60Vdc. The user must
ensure that the incorporation of the unit into the user’s equipment conforms to the relevant sections of BS EN 61010 (Safety Requirements
for Electrical Equipment for Measuring, Control and Laboratory Use).
VARIOUS OPERATING MODES
Specifications liable to change without prior warning
QM 130M Issue 8
September/2001
M.C.
Applies to QM 130M
Measuring a d.c. current.
Rs =
I
1V
5...12Vd.c.
V+
Vin
0V
0V
Load
Rs
Power
Supply
Measuring a voltage in the range
0 to 1V d.c.
0-1Vd.c.
5...12Vd.c.
+
-
V+
Vin
0V
0V
Simultaneous d.c. Voltage and Current Measurement.
Rs =
I
1V
5...12Vd.c.
V+
V+
Vin
Vin
0V
0V
0V
Load
Rs
Rb
Ra
Power
Supply
I
Current
Voltage
V1
V2
+
-
Ra
Rb
Vin
0V
Voltage
Ra=1M
Rb=Ra.
V1-V2
V2
.
+
-
Rs
I
FSD
Vin
0V
Current
Rs=
I
1V
FSD
SCALING
Two external resistors may be used to alter the full scale
reading of the meter - see table for sample values.
Alternatively, use the following formulae to calculate Ra
and Rb. Select the nearest available standard resistor. To
achieve optimum accuracy, use 1% metal film resistors.
Required F.S.R.
Ra
Rb
10V
910k
100k
100V
1M
10k
10mA
0R
100R
100mA
0R
10R