Martel Electronics PTC8010 User Manual
Page 23

23
Command
Description
FAULT?
Returns the error code of an error that has occurred
FUNC?
Returns the current mode of the calibrator
LOCAL
Returns user to manual operation of the calibrator
LOCKOUT
Locks out the keypad of the calibrator, and allows for remote operation only
MEAS
Sets the mode for measuring of the calibrator
OUT
Sets the output of the calibrator
OUT?
Returns the output of the calibrator
REMOTE
Puts the calibrator in remote mode
RTD_TYPE
Sets the RTD type
RTD_TYPE?
Returns the RTD type
RTD_WIRE
Sets the number of wires used by the RTD mode.
RTD_WIRE?
Returns the wire number setting used in the RTD mode
TEMP_UNIT
Sets input/output temperature unit for RTD and TC
TEMP_UNIT?
Returns the temperature unit for RTD and TC
TSENS_TYPE
Sets temperature sensor type.
TSENS_TYPE?
Returns temperature sensor type
VAL?
Returns the measured values
Table 7: Parameter units
Units
Meaning
MV
Voltage in millivolts
V
Voltage in volts
Ohms
Resistance in Ohms
Cel
Temperature in Celsius
Far
Temperature in Fahrenheit
Kel
Temperature in Kelvin
Table 8: Error codes
Error Number
Error Description
100
A non-numeric entry was received where it should be a numeric entry
101
Too many significant digits entered
102
Invalid units or parameter value received
103
Entry is above the upper limit of the allowable range
104
Entry is below the lower limit of the allowable range
105
A required command parameter was missing
107
An invalid CJC_STATE was received
108
An invalid TSENS_TYPE was received