1. After power-up, select the desired output

range. The 200mV, 2V, 20V, and 24mA
ranges can be selected by pressing the
RANGE key until the desired range is
shown on the LCD readout.

2. Observe the proper polarity when

connecting the output jacks. Red is (+)
and Black is (-).

3. Adjust output using the control keys.

When pressed, the keys select a
digit on the LCD readout to be scrolled.
When these keys are pressed and held
down, a blinking cursor appears and
covers up one of the digits indicating which
digit is selected. After a second more, the
cursor starts to move in the direction
selected.

The

v w keys scroll the digit up or

down, respectively. As the outputs are
scrolled, the integral LCD readout displays
the voltage or current present at the output
jacks.

4.

The SP1 or SP2 keys will store or recall a
value when pressed.

To store a value in SP1 or SP2, scroll to the
desired output. Press and hold down the
SP1 or SP2 keys until the word “STORED”
flashes onto the LCD readout.

To recall a value in SP1 or SP2:
momentarily press the SP1 or SP2 keys,
and the value needed from that particular
setpoint should appear on the LCD display.

5.

The voltage output jacks are disabled while
in the current output mode. Likewise, the
current output jacks are disabled when in

3