beautypg.com

KYORITSU 5201 User Manual

Model

background image

ACCESSORIES

The Model 5201 Digital Illumination Meter includes:-

  

⑴ Photocell cover

1each

  

⑵ 9 volt dry battery(006P)

1each

  

⑶ Soft carrying case

1each

  

⑷ Instruction sheet

1each

SPECIFICATIONS

Measurement range

0.1 to 19.990 lux

Range settings

Automatic 3 range switching

Accuracy

±5% of reading and ± 1 digit

Measuring time

3 times per second

Temperature-humidity

characteristics

±3% (at 20℃)

Spectral response

characteristics

Closely related to the spectral luminous

efficiency (of a standard observer).

Angular incident light

characteristics

Closely related to the Lambert's cosine law.

Power source

6F22 battery or equivalent, or DC external

power source.

Operating conditions Temperature

range Humidity range

5℃ to 35℃ 45% to 85%

Dimension

68Ч166Ч32mm

Weight

180 grams (including battery)

FEATURES

The range setting is switched automatically so that you do not require

any switching.

The reading on LCD is held for a preset time (about 20 seconds) so that

you can read the measurement value easily,

Auto Power-Off system is incorporated in the instrument (for

automatically switching off the battery) for a longer battery life. (If the

instrument is used intermittently for about one or two hours per day, it

should be possible to use a fresh battery for about 1000 hours.)

WARNING

⑴ When the measuring button is depressed, “…” should be

displayed.

This will be an indication that the battery is getting weak and should

be replaced.

Use a 9 volt battery and be sure to replace battery with correct

polarities.

⑵ A “- - -“ display will appear when the range is switched. However,

if the display should continue to appear for more than 3 seconds,

this will indicate overranging and measurement are not possible.

⑶ Always remember to put the rubber plug back correctly atter zero

adjustments.

⑷ Do not expose the photocell detector to excessive illumination when

it is not being used, as this will lead to deterioration of the meter

sensitivity. Always remember to cover the photocell detector when

the instrument is stored.

⑸ Keep the photocell detector clean because light transmission will be

obstructed and will be reading errors.

Always wipe the surface clean with a soft and dry cloth, when the

detector is covered with dust and/or dirt.

⑹ Protect the instrument from impact and vibrations as the instrument

could be damaged.

⑺ Do not wipe the plastic instrument body with any chemical solution

such as acetone, ketone, thinner, etc., and do not expose the

instrument to temperature of more than 60℃

⑻ The instrument should be calibrated once a year or so for the

accurate measurement.

Congratulations on your choice of the KYORITSU Digital Illumination Meter Model 5201,a portable and compact photocell illuminance meter

for measuring from 0.1 to 19,990 Lux with auto range switching.

Please read the instructions manual completely before you touch the instrument so that you will be able to use it with greater efficiency over

a longer period of time.

MODEL

5201

DIGITAL ILLUMINATION METER

⑼ Do not open the instrument for repair and calibration exceeding

described in this manual or KYORITSU will not accept the instrument

under the warranty.

Please contact the KYORITSU distributor for instructions when

the instrument seems not be working correctly or when calibration is

required.

⑽ When storing the instrument in the carrying case, put in the manner

illustrated with the measuring button located

on the cover side so that it will be positioned

between the pads on the inside of the cover.

Otherwise, the button could become depressed

(i.e., switched ON), which the battery would be

exhausted.

OPERATIONS

⑴ Take the instrument out of the soft carrying case.

⑵ Put the photocell detector cover by hand completely so that no light

can reach to the detector.

Depress the measuring button for about two seconds.

Check whether the digital display is“00.0”and make sure that an

indication of low battery power “…”is not displayed.

⑶ Take of the photocell detector cover and place the photocell detector

to the surface being measured to receive the incident light.

Then press the measuring button for about 2 seconds.

⑷ When the digital display is stabillzed, release the measuring button.

Reading of the measurement on the display should be held or about

20 seconds for your easy reading and writing memo.

The power will be automatically off after 20 seconds.

⑸ The measuring button has to be depressed while the previous

reading is being held, a new digital display will appear for the

illuminance of the incident light at the time.

⑹ The hand strap on the instrument can be used for setting the

measuring distance. Unhooking one end of the strap and extending it

will work as scale for 30 centimeters.

If the strap is used without unhooking it, the strap can be utilized as

a scale for 15 centimeters.

CAUTON

If the digital display does not become“00.0”in (2) above, zero

adjustment will be required. Detach the rubber plug on the rear surface

of the instrument. While depressing the measuring button, turn the

adjustment knob by using a screwdriver so that the display indicates

00.0”. Put the rubber plug back after adjustments are finished.

Photocell

detector head

Display window

Measuring button

Instrument

Hand strap

Photocell

detector

Rubber

plug

Battery

compartment

15cm

30cm

10-07

92-1345A

INSTRUMENT LAYOUT