― 8 ―

Operating System

:Dual Integration

Display

:Field effect 3・1/2 digit liquid crystal

display with maximum count of 1999

Range selection

:Manual

Overrange Indication

:“1”is displayed on the highest digit

except for 1000A AC range.

Response Time

:Approx. 3 seconds

Sample Rate

:Approx. 3 times per second

Data Hold

:Available on all ranges

Location for use

:Indoor use, Altitude up to 2000m

Temperature & Humidity

:23℃

±

5℃, relative humidity up to 85%

for Specified Accuracy

without condensation

Storage Temperature

:-20∼60℃, relative humidity up to 80%

& Humidity

without condensation

Operating Temperature

:0∼40℃,relative humidity up to 85%

& Humidity

without condensation

Power Source

:One 6 F22 or equivalent battery

Low Battery Warning

:“

”symbol is shown on the display.

Current Consumption

:Approx. 5mA max.(use Back Light:8mA max.)

Standard(Safety)

:IEC61010-1

IEC61010-2-032

CAT.

,AC300V,Pollution degree 2

(EMC)

:EN 61326

・EN 55022

・EN 61000-4-2 (performance criterion B)

・EN 61000-4-3 (performance criterion B)

Overload Protection

:1500A AC max for 1 minute

Withstand Voltage

:3700V AC for 1 minute between

electrical circuit and housing cases

Insulation Resistance

:50MΩ or greater at 1000V between

electrical circuits and housing cases or
metal parts of jaws

Conductor Size

:Approx. 68mm diameter max.

Dimensions/Weight

:250(L)

×

130(W)

×

50(D)mm.

Approx. 600g (including battery)

Accessories

:6F22 battery

Carrying case
Instruction manual

Optional Accessories

:Model 7073 two-way output cord

Peak Hold

:Response time selectable from approx.

10ms and 100ms.
1/

√2 of the peak current is displayed,

which means an RMS reading will be
obtained when current having sinusoidal
waveform in measured.
Peak hold reading varies by approximately

±

1% max of full scale in the first minute at

23

±

5℃ and relative humidity of 75%

without condensation.