beautypg.com

Device description, 6 protection type for gas hazardous areas, 7 temperature classes – KROHNE SMARTSENS PH 8xxx Ex EN User Manual

Page 8: 8 autoclavability

background image

2

DEVICE DESCRIPTION

8

SMARTSENS PH 8XXX

www.krohne.com

07/2014 - 4002951402 - MA SMARTSENS PH 8xxx R02 en

2.6 Protection type for gas hazardous areas

In protection type "intrinsic safety" the sensor meets the requirements of EN 60079-11. The
explosion protection is ensured by limitation of the current and voltage so that no ignitable
energy can occur.
The marking of the sensor is:

II 1G Ex ia IIC T6

II 1G Ex ia IIC T6

II 1G Ex ia IIC T6

II 1G Ex ia IIC T6

2.7 Temperature classes

The permissible temperature range depending on the temperature class of the sensor is:

2.8 Autoclavability

To kill off germs, the sensor may be stored in the autoclave at 140°C / 284°F for 20 minutes.

Components of the Ex marking and their definition

II

IIII

II

Group II explosion protection

1111

Equipment category 1

G

G

G

G

Gas explosion protection

Ex ia

Ex ia

Ex ia

Ex ia

Intrinsically safe equipment, protection level ia

IIC

IIC

IIC

IIC

Gas groups IIA, IIB, IIC approved

T6

T6

T6

T6

Temperature classes T6...T1 approved

Temperature
class

Permissible ambient or product

temperature in [°C]

Permissible ambient or product

temperature in [°F]

T1

-10...+85°C

14...+185°F

T2

-10...+85°C

14...+185°F

T3

-10...+85°C

14...+185°F

T4

-10...+85°C

14...+185°F

T5

-10...+55°C

14...+131°F

T6

-10...+55°C

14...+131°F

The permissible product temperature range can vary depending on the sensor used. Consult the
standard documentation for this information.

INFORMATION!
The sensor is not operated in the autoclave. For more information consult the standard
documentation.

.book Page 8 Monday, July 21, 2014 3:08 PM