beautypg.com

Technical data – KROHNE OPTIFLUX 5000 Flange EN User Manual

Page 25

background image

TECHNICAL DATA

6

25

OPTIFLUX 5000

www.krohne.com

04/2011 - 4000686302 - HB OPTIFLUX 5000 FL R02 en

Inlet run

≥ 5 DN (without disturbing flow, after a single 90° bend)
≥ 10 DN (after a double bend 2x 90°)
≥ 10 DN (behind a control valve)

Outlet run

≥ 2 DN

Dimensions and weights

For detailed information refer to

Dimensions and weights

on page 27.

Materials

Sensor housing

DN15...100: stainless steel AISI 316 (1.4408)

DN150...300: sheet steel (carbon steel)

Measuring tube

ceramic

Connection box

(only remote versions)

Standard: polyurethane coated die-cast aluminium

Option: stainless steel

Grounding rings

Standard:

Standard:

Standard:

Standard:

DN15...100: not included

DN150...300: integrated, stainless steel AISI 316L (1.4404)

Option:

Option:

Option:

Option:

DN15...100: virtual reference (only with IFC 300 converter)

DN150...300: Hastelloy

®

C

Gaskets

DN15...100: PTFE

DN150...300: FPM / FKM O-ring

Measuring electrodes

Standard:

Standard:

Standard:

Standard:

DN15...100: cermet

DN150...300: stainless steel AISI 316 Ti (1.4571)

Option:

Option:

Option:

Option:

DN150...300: platinum, Hastelloy

®

C, low noise (basis Hastelloy C4,

tantalum, titanium)

Process connections

EN 1092-1

DN200...300: PN10

DN100..150: PN16

DN15...80: PN40

ASME

Standard:

Standard:

Standard:

Standard:

1...12": 150 lb

½

": 300 lb

Option:

Option:

Option:

Option:

1...3": 300 lb

Electrical connections

Signal cable

Only for remote systems.

Type A

Standard cable, double shielded.

Max. length: 600 m / 1950 ft (dep. on electrical conductivity and measuring

sensor). See documentation of the converter for more information.

Type B

Optional cable, triple shielded.

Max. length: 600 m / 1950 ft (dep. on electrical conductivity and measuring

sensor). See documentation of the converter for more information.