beautypg.com

Installation, 5 mounting, 1 torques and pressures – KROHNE OPTIFLUX 5000 Flange EN User Manual

Page 16

background image

3

INSTALLATION

16

OPTIFLUX 5000

www.krohne.com

04/2011 - 4000686302 - HB OPTIFLUX 5000 FL R02 en

3.5 Mounting

3.5.1 Torques and pressures

DN15...100
Tighten the bolts in fixed order, see picture:

• Step 1: by hand
• Step 2: approx. 25% of max. torque
• Step 3: approx. 50% of max. torque
• Step 4: approx. 80% of max. torque
• Step 5: 100% of max. torque given in table

EN 1092-1

WARNING!
Please use "not lubricated" Stainless steel A2 / 6.9 class bolts.

INFORMATION!
Diameters DN80 and DN100 have 8 holes per flange, please continue in the same way to tighten
the other bolts.

CAUTION!
With the instrument, 4 PTFE gaskets are delivered (2 to be used with installation, 2 as spare). No
other gaskets are required.

Nominal size

DN [mm]

Pressure

rating

Bolts

Max. torque

[Nm]

15

PN 40

4 × M 12

73.5

25

PN 40

4 × M 12

73.5

40

PN 40

4 × M 16

178

50

PN 40

4 × M 16

178

80

PN 40

8 × M 16

178

100

PN 16

8 × M 16

178