ALTOFLUX 2W IFM 4042 K - EEx

9

2.5 Safety-technical

data

Important

The functional-technical data must also be regarded,

therefore see the standard installation and operating instructions.


Safety-technical data of output circuits
Terminal

Function

Electrical data (per circuit)

designation

Type of protection "EEx ib" Type of protection "EEx e"

Circuit 1
I, I

Current output, passive
(2-wire connection)
4 - 20 mA, HART =
possible

U

n

= 14 - 36 Vdc

I

n

= 4 - 20 mA

U

m

= 250 V

Circuit 2
1L=, 0L=

Additional power supply
or Power Booster,
(4-wire connection)
additional to circuit 1
(option)

Maximum values:
U

i

= 30 V, I

i

= 100 mA,

P

i

= 1.0 W

Ci

= 20 nF, Li = 0

U

m

= 60 V

U

n

= 14 - 36 Vdc

I

n

= 22 mA

U

m

= 250 V

Circuit 3
B1, B

B2, B

Passive pulse/status
output 1
Passive pulse/status
output 2

Maximum values:
U

i

= 30 V, Ii = 100 mA,

P

i

= 1.0 W

C

i

= 0, Li = 0

U

m

= 60 V

Maximum values:
U

= 36 V

I

= 100 mA

U

m

= 250 V

2.6 Connection

examples

In the following sections examples for connection of the IFM 4042 K-EEx compact flowmeter are
described for operation in the 2-wire mode as well as in the 2x2-wire mode.

2.6.1 Example of IFM 4042 K in 2-wire mode
The diagram shows an IFM 4042 K-EEx with the terminal compartment in version A (EEx de [ib]
with Um = 60 V). The flowmeter is connected through a transmitter power supply ("EEx i"
approved) in 2-wire mode. If data communication with the flowmeter through the HART protocol is
required, the transmitter power supply unit must be HART compatible. Terminals I, I

⊥ are not

polarity sensitive.
The entity parameters of the "EEx i" approved transmitter power supply, including the cable
capacitances and inductances, must fit the entity parameters of the IFM 4042 K-EEx-EEx compact
flowmeter, namely U

i

= 30 V, I

i

= 100 mA, C

i

= 200 nF, L

i

= 0. Suitable HART compatible trans-

mitter power supplies that can be used in combination of the Altoflux 2W IFM 4042 K-EEx are:

• Phoenix

PI/Ex-ME-RPSS-I/I

• CEAG

6/420

IFM 4042 K-EEx

I

I

HTT

4-20 mA

4-20 mA

Transmitter Power

Supply (EEx i)

Vs

Hazardous Area

Process/

Display Unit

Safe Area