DESCRIPTION OF HART INTERFACE
36
SMARTSENS PH 8510
www.krohne.com
11/2013 - 4002544402 - MA SMARTSENS PH 8510 R02 en
7.4 Inputs/outputs and HART
®
dynamic variables and device variables
Code = device variable code
Device variables
The HART
®
dynamic variable PV is always connected to the HART
®
current output which is
assigned to the pH value.
7.5 Field Communicator 475 (FC 475)
The Field Communicator is a hand terminal from Emerson Process Management that is
designed to configure HART
®
and Foundation Fieldbus devices. Device Descriptions (DDs) are
used to integrate different devices into the Field Communicator.
7.5.1 Installation
The HART
®
Device Description for the sensor must be installed on the Field Communicator.
Otherwise only the functions of a generic DD are available to the user and the entire device
control is not possible. A "Field Communicator Easy Upgrade Programming Utility" is required to
install the DDs on the Field Communicator.
The Field Communicator must be equipped with a system card with "Easy Upgrade Option". For
details consult the Field Communicator User’s Manual.
HART
®
dynamic variable
PV
SV
TV
QV
pH
Electrode voltage
Temperature
Resistance
HART
®
device variable
Code
Type
pH value
0
linear
Electrode voltage
1
linear
Temperature
2
linear
Resistance
3
linear