beautypg.com

Operation, 3 status messages and diagnostic information – KROHNE SMARTSENS PH 8510 EN User Manual

Page 23

background image

OPERATION

4

23

SMARTSENS PH 8510

www.krohne.com

11/2013 - 4002544402 - MA SMARTSENS PH 8510 R02 en

4.3 Status messages and diagnostic information

Measurements out of specification

Failure

Maintenance

Message

Message

Message

Message

Description

Description

Description

Description

Action

Action

Action

Action

S01

pH value > pH max
pH value out of range

Consider the measuring range limits, otherwise
select a suitable sensor for the process conditions of
the application.

S02

pH value < pH min
pH value out of range

S03

Temp. value > Temp max
Temperature value out of range

S04

Temp. value < Temp min
Temperature value out of range

Message

Message

Message

Message

Description

Description

Description

Description

Action

Action

Action

Action

F01

Glass resistance > max. MOhm

Replace the sensor

F02

Glass resistance < min. MOhm

Message

Message

Message

Message

Description

Description

Description

Description

Action

Action

Action

Action

M01

Slope < -65 mV/pH

Recalibrate the sensor. For more information refer
to

Calibration with PACTwareTM on page 17

M02

Slope > -50 mV/pH

M03

Zero point < -99 mV

M04

Zero point > +99 mV

M05

Maintenance interval expired

.book Page 23 Monday, November 18, 2013 2:46 PM