Overflow protection

(max./A-operation)

Dry run protection

(min./B-operation)

λ

sd

9 FIT

9 FIT

λ

su

203 FIT

202 FIT

λ

dd

9 FIT

9 FIT

λ

du

43 FIT

45 FIT

MTBF = MTTF + MTTR

3.79 x 10

6

h

3.79 x 10

6

h

Fault reaction time

<

1.5 sec.

Single channel architecture (1oo1D)

SIL

SIL

2

HFT

0

Sensor type

Type A

Overflow protection

(max./A-operation)

Dry run protection

(min./B-operation)

SFF

83 %

83 %

PFD

avg

T

Proof

=

1 year

T

Proof

=

5 years

T

Proof

=

10 years

<

0.019 x 10

-2

<

0.095 x 10

-2

<

0.189 x 10

-2

<

0.020 x 10

-2

<

0.097 x 10

-2

<

0.195 x 10

-2

PFH

<

0.043 x 10

-6

/h

<

0.045 x 10

-6

/h

The chronological sequence of PFD

avg

is nearly linear to the operating

time over a period up to 10 years. The above values apply only to the
T

Proof

interval after which a recurring function test must be carried out.

Failure rates

Fault reaction time

Specific characteristics

Time-dependent proc-
ess of PFD

avg

OPTISWITCH series 5000 • - NAMUR

9