4 Mounting

4.1 Mounting procedure

The numbers in brackets refer to the illustrations on the

following pages.

1

2

Fig. 1: Lock fitting ARV 60 - unpressurised

1

Lock fitting

2

Terminal screws (3 pcs.)

1 Screw the lock fitting (1) with a resistant seal ring into the

thread of your vessel and tighten the lock fitting (1) on the

hexagon

2 Clean the connection tube of the sensor and the lock fitting

carefully and remove grease, oil and dirt. Insert the sensor

into the lock fitting. Slide the tube into the requested

position and hold it

3 Make sure that the sensor is in the correct position (height.

The height adjustment of the sensor determines also the

switching point

4 Tighten the terminal screws (2) with a torque of 4 ±1 Nm

(3 ±0.7 lbf ft)
The terminal screws (2) press lightly into the tube and fix

the tube of the sensor in this position.

6

Lock fitting ARV 60 - unpressurised for OPTISWITCH 5200

Mounting