Page 23

background image

l

Wrong characteristics selected

à

Set the correct characteristics on the characteristics

reversal switch (overfill protection; dry run protection).

Wiring should be carried out according to the quiescent

current principle.

?

Signal lamp flashes red

l

Fault in the electronics

à

Exchanging the electronics module

?

Signal lamp flashes red

l

instrument defective

à

Exchange instrument or return instrument for repair

7.3 Instrument repair

If a repair is necessary, please proceed as follows:
You can download a return form from our website http://www.

krohne-mar.com/fileadmin/media-lounge/PDF-Download/

Specimen_e.pdf.
By doing this you help us carry out the repair quickly and

without having to call back for needed information.

l

Print and fill out one form per instrument

l

Clean the instrument and pack it damage-proof

l

Attach the completed form and possibly also a safety data

sheet to the instrument

OPTISWITCH 5200 C, 5250 C - with NAMUR output

23

Maintenance and fault rectification