Single-colour control lamp (LED) for indication of the switching

condition

l

red = High current >= 2.2 mA

l

dark = Low current <=1 mA

l

red (flashing) = Failure <=1 mA

The characteristics reversal can be carried out with the DIL

switch. You can choose between falling characteristic curve

(switch position max.) and rising characteristic curve (switch

position min.). This allows you to output the desired current.
Modes

l

min. - rising characteristic curve (High current when

immersed)

l

max. - falling characteristics (Low current when immersed)

The NAMUR output can be switched over to falling or rising

characteristics (see also "Function chart").
For applications acc. to WHG, the DIL switch must be set to

position max.

With this DIL switch (3) you can set the switching point to

liquids having a density between 0.5 and 0.7 g/cm³ (0.018 and

0.025 lbs/in³). With the basic setting, liquids with a density of

>0.7 g/cm³ (>0.025 lbs/in³) can be detected. In liquids with

lower density, you must set the switch to >0.5 g/cm³

(>0.018 lbs/in³). The specifications for the position of the

switching point relate to water - density value 1.0 g/cm³

(0.036 lbs/in³). In products with a different density, the

switching point will shift in the direction of the housing or tuning

fork end depending on the density and type of installation.

The simulation key is located in a recess on the upper side of

the oscillator. Push the simulation key with a suitable object

(screwdriver, pen, etc.).
When the key is pushed, a line break between sensor and

processing unit is simulated. The signal lamp on the sensor

extinguishes. The measuring system must signal a fault and

take on a safe condition when the key is pushed.
Keep in mind that the connected (downstream) instruments will

be activated during operation. This allows you to check the

correct function of the measuring system.

Signal lamp (1)

Characteristics reversal (2)

Sensitivity adjustment (3)

Simulation key (4)

OPTISWITCH 5100 C, 5150 C - with NAMUR output

19

Set up