on correct function. The following operating conditions are

simulated during the test:

l

Fault signal

l

Empty signal

l

Full signal

Check if all three switching conditions occur in the correct

sequence and the stated time period. If this is not the case,

there is a fault in the measuring system (see also the operating

instructions manual of the signal conditioning instrument).

Keep in mind that connected instruments are activated during

the function test. By doing this, you can check the correct

function of the measuring system.

(after releasing the test key)

Sensor

current -

Sensor

Level re-

lay A -

overfill

protection

Level re-

lay B - dry

run pro-

tection

Signal

lamp A -

Overfill

protection

Signal

lamp B -

Dry run

protection

Fail safe

relay

Control

lamp

1. Fault signal

(approx. 3 s)

<2.0 mA

currentless currentless

currentless

2. Empty signal

(approx. 1.5 s)

approx. 8

mA

energized

currentless

energized

3. Full signal

(approx. 1.5 s)

approx. 16

mA

currentless energized

energized

4. Return to the actual

operating condition

Note:

Mode B is not permissible when used as part of an overfill

protection system acc. to WHG and use in measuring chains

acc. to IEC 61508.

You can carry out the function test with the outputted current

values also directly with a PLC or a process control system

(WHG).

Test procedure

OPTISWITCH 5100 C, 5150 C - with two-wire output

23

Set up

30430

-EN

-060831