1 Unscrew the housing cover
2 Loosen compression nut of the cable entry
3 Remove approx. 10 cm (4 in) of the cable mantle, strip

approx. 1 cm (0.4 in) insulation from the ends of the

individual wires

4 Insert the cable into the sensor through the cable entry
5 Lift the opening levers of the terminals with a screwdriver

(see following illustration)

6 Insert the wire ends into the open terminals according to

the wiring plan

7 Press the opening lever of the terminals downward, you will

hear the terminal spring closing

8 Check the hold of the wires in the terminals by lightly

pulling on them

9 Tighten the compression nut of the cable entry, the seal

ring must completely encircle the cable

10 If necessary, carry out a fresh adjustment
11 Screw the housing cover back on
The electrical connection is hence finished.

Fig. 8: Connection steps 5 and 6

5.3 Wiring plan, single chamber housing

The following illustrations apply to the non-Ex as well as to the

EEx d version.

16

OPTISWITCH 3100 C - with relay output

Connecting to voltage supply

29953

-EN
-060830