beautypg.com

3 operation, 4 storage and transport – KROHNE OPTISWITCH 3000C Contactless EN User Manual

Page 7

background image

Fault monitoring

The electronics of OPTISWITCH 3000 C continuously mon-

itors the following criteria:

l

Correct vibrating frequency

l

Line break to the piezo drive

If one of the stated malfunctions is detected or in case of

power failure, the electronics takes on a defined switching

condition, i.e. the contactless electronic switch opens (safe

condition).

The tuning fork is piezoelectrically energised and vibrates at its

mechanical resonance frequency of approx. 150 Hz. When the

tuning fork is submerged in the product, the vibration

amplitude changes. This change is detected by the integrated

oscillator and converted into a switching command.

OPTISWITCH 3000 C is a compact instrument, i.e. it can be

operated without external evaluation system. The integrated

electronics evaluates the level signal and outputs a switching

signal. With this switching signal, a connected device can be

directly activated (e.g. a warning system, a PLC, a pump etc.).
The exact range of the power supply is stated in the "Technical

data" in the "Supplement".

3.3 Operation

OPTISWITCH 3000 C can detect products with a density

>0.08 g/cm³ (>0.003 lbs/in³).
On the electronics module you will find the following indicating

and adjustment elements:

l

signal lamp for indication of the switching condition (green/

red)

l

Mode switch for selecting the switching condition (min./

max.)

3.4 Storage and transport

Your instrument was protected by packaging during transport.

Its capacity to handle normal loads during transport is assured

by a test according to DIN 55439.

Physical principle

Power supply

Packaging

OPTISWITCH 3000 C - with contactless electronic switch

7

Product description

30058

-EN

-060830