beautypg.com

Electrical connections, 4 laying electrical cables correctly – KROHNE OPTISONIC 3400 Quickstart EN User Manual

Page 22

background image

3

ELECTRICAL CONNECTIONS

22

OPTISONIC 3400

www.krohne.com

11/2013 - QS OPTISONIC 3400_4002741202_R02-en

100…230 VAC

• Connect the protective ground conductor PE of the mains power supply to the separate

terminal in the terminal compartment of the signal converter.

• Connect the live conductor to the L terminal and the neutral conductor to the N terminal.

24 VAC/DC

• Connect a functional ground FE to the separate U-clamp terminal in the terminal

compartment of the signal converter.

• When connecting to functional extra-low voltages, provide a facility for protective separation

(PELV) (VDE 0100 / VDE 0106 and/or IEC 364 / IEC 536 or relevant national regulations).

3.4 Laying electrical cables correctly

1 Lay the cable in a loop just before the housing.
2 Tighten the screw connection of the cable entry securely.
3 Never mount the housing with the cable entries facing upwards.
4 Seal cable entries that are not needed with a plug.

Figure 3-5: Protect housing from dust and water