SAFETY-RELATED CHARACTERISTICS

12

H250 M40

www.krohne.com

06/2013 - 4000904202 MA H250 M40 SIL R02 en

7.2 Specific safety-related characteristics

Under the assumptions described in 7.1 and the definitions given in section 4 the following tables
show the failure rates according to IEC 61508:

H250/M40/K*-SK

H250/M40/K*-SK

H250/M40/K*-SK

H250/M40/K*-SK with 1 or 2 fail-safe limit switches

fail-safe limit switches

fail-safe limit switches

fail-safe limit switches (MIN/MAX) 1 and welded

welded

welded

welded process

connections

H250/M40/K*-SK

H250/M40/K*-SK

H250/M40/K*-SK

H250/M40/K*-SK with 1 or 2 standard limit switches

standard limit switches

standard limit switches

standard limit switches (MIN/MAX) 6 and welded

welded

welded

welded process

connections

λ

SD

λ

SU

λ

DD

λ

DU

SFF 2 DC

D

MTBF

SIL AC 3

Stress profile 2
(low stress)

0 FIT

50 FIT

10 FIT

38 FIT

61%

20%

487 years

SIL2

Stress profile 4
(high stress)

0 FIT

50 FIT

10 FIT

73 FIT

45%

12%

356 years

SIL1

T[Proof] 4

1 year

2 years

5 years

PFD

AVG

5

1.82*E-04

3.46*E-04

8.41*E-04

λ

SD

λ

SU

λ

DD

λ

DU

SFF 2 DC

D

MTBF

SIL AC 3

Stress profile 2
(low stress)

0 FIT

81 FIT

10 FIT

77 FIT

54%

11%

376 years

SIL1

Stress profile 4
(high stress)

0 FIT

81 FIT

10 FIT

112 FIT

44%

8%

293 years

SIL1

T[Proof] 4

1 year

2 years

5 years

PFD

AVG

5

3.68*E-04

7.02*E-04

1.70*E-03