Device description – KROHNE H250 M9S ATEX II3D Ex t-II3G Ex nA EN User Manual

Page 9

background image

DEVICE DESCRIPTION

2

9

H250 M9

www.krohne.com

10/2012 - 4000646002 MA H250/M9-Ex-II3GD-AD-R06-en

Using a heat resistant connecting cable

1 Maximum value of the product temperature for the use of a standard connecting cable. For
higher product temperatures a connecting cable with a temperature resistance of 100°C / 212°F
is required.

HT version - high-temperature version with projecting indicator

The permitted ambient temperature range is indicated on the nameplate; depending on the
device version it is T

amb

= -40...+60°C / -40...+140°F or T

amb

= -25...+60°C / -13...+140°F.

Temperature table in °C

Heating jacket

T

m

1

none

with

HT-version

T

amb

≤ 40

T

amb

≤ 60

DN15, DN25

DN15

-

150

x

-

236

DN50

DN25

-

127

x

-

171

DN80, DN100

DN50, DN80

-

109

x

-

145

Temperature table in °F

Heating jacket

T

m

1

none

with

HT-version

T

amb

≤ 104

T

amb

≤ 140

DN15, DN25

DN15

-

302

x

-

456

DN50

DN25

-

260

x

-

340

DN80, DN100

DN50, DN80

-

228

x

-

293