2.8 Electrical data

Built-in equipment for the variable area flowmeter may only be connected to separate
intrinsically safe circuits with the following maximum values:

Electrical equipment

Nominal voltage

Nominal current

Limit switch K1 / K2

8 VDC

1...3 mA

Signal output ESK II and ESK2A

24 VDC ± 25%

4...20 mA with HART

®

communication

ESK3-PA Profibus transmitter

1

9...24 VDC

12 mA

1 Further information and instructions for operation of the ESK3-PA Profibus transmitter are provided in separate sup-

plementary instructions.

Maximum values

Built-in equipment

U

i

[V]

Ii[mA]

P

i

[mW]

C

i

[nF]

L

i

[μH]

ESKII / ESK2A

30

100

1000

20

~ 0

ESK3-PA

1

24

380

5320

~ 0

~ 0

I7S23,5-N

SC3,5-N0

16

25

64

150

150

16

52

169

150

150

SJ3.5-SN

SJ3.5-S1N

16

25

64

30

100

16

52

169

30

100

1 FISCO Field device