Device description – KROHNE H250 M40 ATEX II2D Ex t-II2G Ex d EN User Manual

Page 10

background image

2

DEVICE DESCRIPTION

10

H250 M40

www.krohne.com

07/2012 - 4002069601 - MA H250/M40-Ex-II2GD-AD R01

Reference point observation

Reference point observation

Reference point observation

Reference point observation

The permissible product and ambient temperatures may be exceeded or undershot as long as
the permissible temperature range of the reference point of the display is not exceeded. The
following table contains the permissible maximum values at the reference point. Note:

• Reference point is the connection of the equipotential bonding conductor of the M40 indicator.
• The temperatures at the reference point are to be determined in the most unfavourable

operating situation.

• The required measuring uncertainty is max. 2K.
• Measuring conditions and results are to be permanently recorded in a suitable format.
• The measuring unit should be insulated properly.

Maximum permitted temperatures at reference point

Maximum permitted temperature at the reference
point T

Ref

[°C]

Heating
jacket

TK

T6...T1

T6

T5...T1

none

with

T

Connecting cable

[°C] standard (70°C)

Heat resistant (90°C)

DN15

DN25

DN50

DN80

DN100

DN15

DN25

DN50

DN80

64

74

84

Maximum permitted temperature at the reference
point T

Ref

[°F]

Heating
jacket

TK

T6...T1

T6

T5...T1

none

with

T

Connecting cable

[°F] standard (158°F) Heat resistant (194 °F)

DN15

DN25

DN50

DN80

DN100

DN15

DN25

DN50

DN80

147

165

183

AD_H250_M40_II2GD_R01_en_4002069601_PRT.book Page 10 Friday, July 27, 2012 8:17 AM