beautypg.com

Modbus protocol, 10 calibration procedures – KROHNE IFC 300 Converter Modbus EN User Manual

Page 18

background image

6

MODBUS PROTOCOL

18

IFC 300

www.krohne.com

07/2010 - 4000410703 - AD Modbus IFC 300 R03 en

6.10 Calibration procedures

6.10.1 Zero Flow Calibration

1

User reduces the actual flow through the meter to

User reduces the actual flow through the meter to

User reduces the actual flow through the meter to

User reduces the actual flow through the meter to
zero

zero

zero

zero

2

Start zero calibration:

Start zero calibration:

Start zero calibration:

Start zero calibration:
Set single coil 2000 to 1 (Modbus Fct. 0x05)

3

Check status of calibration

Check status of calibration

Check status of calibration

Check status of calibration
Read single coil 2000 (Modbus Fct. 0x01)

3a

[2000 == 1: calibration running]

3b

[2000 == 0: calibration complete]

4

Read Calibration Value

Read Calibration Value

Read Calibration Value

Read Calibration Value
Read Input Register 20000 type float
(Modbus Fct. 0x04)

5

User checks the calibration value:

User checks the calibration value:

User checks the calibration value:

User checks the calibration value:
-0.01...+0.01: good results
-0.1...+0.1: acceptable result for difficult
application

5a

Not ok

5b

ok

6

User tries to improve the calibration situation

User tries to improve the calibration situation

User tries to improve the calibration situation

User tries to improve the calibration situation

6a

Improvement done

6b

Improvement not possible
-Break-

7

Write the calibration value

Write the calibration value

Write the calibration value

Write the calibration value
Write Holding Register 43024 type float
(Modbus Fct. 0x10)

8

Activate new values

Activate new values

Activate new values

Activate new values
Set single coil 1001 to 1 (Modbus Fct. 0x05)

8a

This results in a warm start of the device.
Will take maximum 5s.

.book Page 18 Friday, July 9, 2010 10:31 AM