OPERATION
2
9
USB ADAPTER PLUS
www.krohne.com
06/2014 - 4003748601 - QS USB_Adapter_Plus R03 en
Menu C2: Inputs/outputs (I/O's)
C1.2 Filter
C1.2
Filter
C1.2.1
Min. limitation
-100...-0,001 (m/s)
The output freezes at this value.
C1.2.2
Max. limitation
0,001...100 (m/s)
The output freezes at this value.
C1.2.3
Flow direction
Normal direction (default), Reverse direction
C1.2.4
Time constant
0...100 (s)
Manual input
C1.2.5
Low flow cut off threshold
0...10 (m/s)
Manual input
C1.2.6
Low flow cut off
hysteresis
0...10 (m/s)
Manual input
C1.3 Information
C1.3
Information
No.
Function / subfunction
Settings / descriptions
C2.1 Hardware
C2.1.0
Hardware
Frequency output, Pulse output
C2.1.1
Terminals D
Program frequency output in submenu 2.2.0.1.
Program pulse output in submenu 2.2.0.2.
C2.2 Frequency output (if selected in 2.1.0 "hardware")
C2.2.1
Pulse shape
Symmetric, Fixed, Automatic (default)
C2.2.2
Pulse width
5E-5...2 (s)
C2.2.3
100 % pulse rate
0,01…10000 (Hz)
C2.2.4
Range min.
-1E
+30
...+1E
+30
(m
3
/s)
Manual input.
C2.2.5
Range max.
-1E
+30
...+1E
+30
(m
3
/s)
Manual input.
C2.2.6
Polarity
Both polarity, Positive polarity (default), Negative polarity, Absolute polarity
C2.2.7
Limitation min.
-1E
+30
...+1E
+30
(%)
C2.2.8
Limitation max.
-1E
+30
...+1E
+30
(%)
C2.2.9
Low flow cut off threshold
0...20 (%)
Manual input.
C2.2.10
Low flow cut off
hysteresis
0...20 (%)
Manual input.
C2.2.11
Time constant
0...100 (s)
C2.2.12
Invert signal
Off (default), On