Menu C2: Inputs/outputs (I/O's)

C1.2 Filter

C1.2

Filter

C1.2.1

Min. limitation

-100...-0,001 (m/s)

The output freezes at this value.

C1.2.2

Max. limitation

0,001...100 (m/s)

The output freezes at this value.

C1.2.3

Flow direction

Normal direction (default), Reverse direction

C1.2.4

Time constant

0...100 (s)

Manual input

C1.2.5

Low flow cut off threshold

0...10 (m/s)

Manual input

C1.2.6

Low flow cut off

hysteresis

0...10 (m/s)

Manual input

C1.3 Information

C1.3

Information

No.

Function / subfunction

Settings / descriptions

C2.1 Hardware

C2.1.0

Hardware

Frequency output, Pulse output

C2.1.1

Terminals D

Program frequency output in submenu 2.2.0.1.

Program pulse output in submenu 2.2.0.2.

C2.2 Frequency output (if selected in 2.1.0 "hardware")

C2.2.1

Pulse shape

Symmetric, Fixed, Automatic (default)

C2.2.2

Pulse width

5E-5...2 (s)

C2.2.3

100 % pulse rate

0,01…10000 (Hz)

C2.2.4

Range min.

-1E

+30

...+1E

+30

(m

3

/s)

Manual input.

C2.2.5

Range max.

-1E

+30

...+1E

+30

(m

3

/s)

Manual input.

C2.2.6

Polarity

Both polarity, Positive polarity (default), Negative polarity, Absolute polarity

C2.2.7

Limitation min.

-1E

+30

...+1E

+30

(%)

C2.2.8

Limitation max.

-1E

+30

...+1E

+30

(%)

C2.2.9

Low flow cut off threshold

0...20 (%)

Manual input.

C2.2.10

Low flow cut off

hysteresis

0...20 (%)

Manual input.

C2.2.11

Time constant

0...100 (s)

C2.2.12

Invert signal

Off (default), On