Menu C1: Process input

No.

Function / subfunction

Settings / descriptions

C1.1 Calibration

C1.1

Calibration

C1.1.1

Zero calibration

Measured zero point (m/s).

Use the "calibration" button to calculate a new zero. Calibration procedure

takes 30 seconds.

C1.1.2

Size entry selection

Value (mm), Table (mm, inch)

For "Value": manual input of flow meter size in box 1.1.3.

For "Table": choose from pull down menu in box 1.1.4.

C1.1.5

GK selection

GK, GKL

For "GK": value is shown in box 1.1.6.

For "GKL": value is shown in box 1.1.7

C1.1.8

Coil resistance

10...220 Ω

Configured, can be changed.

C1.1.9

Field frequency

1/6...2 * line frequency

Choose from pull down menu.

1.1.10

Select settling

Standard (predefined), Manual

1.1.11

Settling time

0,001...0,25 (s)

Manual input

1.1.12

Line frequency

50 or 60 Hz

1.1.13

Measurement conductivity Off, On

Off: Default

1.1.14

Number values

0...10

Number of measurements taken to calculate the average conductivity (if

1.1.13 is switched on).