INSTALLATION

10

OPTIFLUX 4000

www.krohne.com

04/2010 - 7309852300 - QS OPTIFLUX 4000 R03 en

2.5.11 Pump

2.5.12 Temperatures

Figure 2-14: Installation after pump

CAUTION!
Protect the device from direct sunlight.

Temperature range

Process [°C]

Ambient [°C]

Process [°F]

Ambient [°F]

min.

max.

min.

max.

min.

max.

min.

max.

PTFE

Separate flow sensor

-40

180

-40

65

-40

356

-40

149

Compact with IFC 300

-40

140

-40

65

-40

284

-40

149

Compact with IFC 100

-40

140

-40

65

-40

284

-40

149

PFA

Separate flow sensor

-40

180

-40

65

-40

356

-40

149

Compact with IFC 300

-40

140

-40

65

-40

284

-40

149

Compact with IFC 100

-40

140

-40

65

-40

284

-40

149

ETFE

Separate flow sensor

-40

120

-40

65

-40

248

-40

149

Compact with IFC 300

-40

120

-40

65

-40

248

-40

149

Compact with IFC 100

-40

120

-40

65

-40

248

-40

149

Hard rubber

Separate flow sensor

1

-5

80

-40

65

23

176

-40

149

Compact with IFC 300

1

-5

80

-40

65

23

176

-40

149

Compact with IFC 100

1

-5

80

-40

65

23

176

-40

149

PU

Separate flow sensor

-5

65

-40

65

23

149

-40

149

Compact with IFC 300

-5

65

-40

65

23

149

-40

149

Compact with IFC 100

-5

65

-40

65

23

149

-40

149

1 Hard rubber liner is available for Ex-versions only