Jenway 73 Series User Manual
Page 6

6
1.2.3.3 Sipper pump
For this accessory as well as removing the passive 
accessory base plate, the front panel of the instrument 
must also be removed. Loosen screws 5 and 6 until the 
front panel can be lifted out in a forward direction. 
Take the sipper base plate. Connect the power supply 
in the bottom of the sample chamber to the connector 
on the underside of the base plate. Place the base plate 
in the sample chamber. Replace screws 1 to 4. Take the 
sipper front panel and slot into place before re-tightening 
screws 5 and 6. 
The tubing should be connected depending on the 
function that the sipper pump is going to perform. All 
tubing must be kept as short as possible and the tubing 
must not be allowed to obstruct the ligth path. 
For sipping:
1. Connect the sipper pump tubing to the outlet port on 
the flow-through cuvette.
2. Secure the tubing using the clip on the righthand side 
of the pump head.
3. Ease the tubing round the rollers by carefully rotating 
them clockwise, by hand. Clamp the tubing into the clip 
on the left hand side of the motor.
4. Once secured, ensure the tubing is routed into the two 
retaining clips located on the base plate at the side of the 
pump head.
5. Cut the tubing at the point where it fits comfortably 
onto the left hand tube located on the inside of the front 
bulk head. 
6. Connect a suitable length of this tubing to the external 
waste pipe.
Bi directional flow A 
(sipping)
Continuous flow
1.
7.
3.
4.
9.
6.
5.
8.
2.
Bi directional flow B 
(pumping)
5.
6.
