Technical specifications – HT instruments HT307 User Manual
Page 8
HT307
EN - 7
6. TECHNICAL SPECIFICATIONS
6.1. CHARACTERISTICS
Accuracy is referred to the following reference conditions: 23°C ± 5°C with RH < 70%.
The luxmeter is calibrated with a standard incandescent lamp at color temperature
2856°K.
Measuring range and accuracy
Range in lux
20 200 2000 20k 200k
Resolution
0.01 0.1 1 10 100
Accuracy
± (3% rdg)
Range in fc
20 200
2000 20k
Resolution
0.01 0.1 1 10
Accuracy
± (3% rdg)
NOTE: 1fc=10.76Lux,
1Klux=1000Lux, 1Kfc=1000fc
for temperature/color different from the reference, accuracy is 6% rdg
Accuracy on angular deviation from cosine characteristic
30° ±
2%
60° ±
6%
80° ±
25%
6.1.1. Spectral
sensitivity
The photodiode with filters makes the spectral sensitivity meet the CIE photo-optic curve
V(
λ) as shown below.