Technical specifications – HT instruments HT307 User Manual

Page 8

background image

HT307

EN - 7

6. TECHNICAL SPECIFICATIONS

6.1. CHARACTERISTICS
Accuracy is referred to the following reference conditions: 23°C ± 5°C with RH < 70%.
The luxmeter is calibrated with a standard incandescent lamp at color temperature
2856°K.

Measuring range and accuracy

Range in lux

20 200 2000 20k 200k

Resolution

0.01 0.1 1 10 100

Accuracy

± (3% rdg)

Range in fc

20 200

2000 20k

Resolution

0.01 0.1 1 10

Accuracy

± (3% rdg)

NOTE: 1fc=10.76Lux,

1Klux=1000Lux, 1Kfc=1000fc

for temperature/color different from the reference, accuracy is 6% rdg

Accuracy on angular deviation from cosine characteristic

30° ±

2%

60° ±

6%

80° ±

25%


6.1.1. Spectral

sensitivity

The photodiode with filters makes the spectral sensitivity meet the CIE photo-optic curve
V(

λ) as shown below.

0

10

20

30

40

50

60

70

80

90

100

380

440

500

560

620

680

740

V(λ)
HT307


Fig. 2: CIE photo-optic curve V(

λ).


6.1.2. Photo

detector

The photo detector is a silicon photodiode with spectral response filter.

Relative sensitivit

y

(

%

)

Wavelength (nm)