beautypg.com

18 s, Oftware – Detcon FP-624C User Manual

Page 26

background image

3.18 S

OFTWARE

F

LOW

C

HAR

Model FP-624C Combustible Gas Sensor PG.26

NORMAL

OPERATION

CALIBRATION

PGM1 (3)

PGM2 (3)

PGM1 (M) PGM2 (M)

PGM1 (3)

SET ALM1 @ ##%

INC

DEC

PGM1 (3)

PGM2 (M)

1-ZERO 2-SPAN

LEGEND

PGM1 - program switch location #1

PGM2 - program switch location #2

(M) - momentary pass of magnet

(3) - 3 second hold of magnet

(30) - 30 second hold of magnet

INC - increase

DEC - decrease

# - numeric value

AUTO ZERO

AUTO SPAN

SET ALARM 1 LEVEL

LEL 0-100 V#.#

PGM1 (3)

PGM2 (M)

VIEW PROG STATUS

ALM1 SET @ ##%

ALM2 SET @ ##%

GAS FACTOR #.##

PGM1 (M) PGM2 (M)

PGM1 (3)

SET ALM2 @ ##%

INC

DEC

PGM1 (3)

PGM2 (M)

SET ALARM 2 LEVEL

PGM2 (3)

PGM2 (3)

PGM2 (3)

CAL LEVEL @ ##%

ALM1 (Firing Direction)

ALM1 (Latch State)

ALM1 (Energize State)

ALM2 (Firing Direction)

ALM2 (Latch State)

ALM2 (Energize State)

FLT (Latch State)

FLT (Energize State)

PGM1 (3)

PGM1 (M)

PGM2 (30) PGM2 (M)

ALARM RESET

ALARM RESET

485 ID SET @ #

SENSOR LIFE ##%

PGM1 (M) PGM2 (M)

PGM1 (3)

INC

DEC

PGM1 (3)

PGM2 (M)

PGM1 (M) PGM2 (M)

PGM1 (3)

CAL FACTOR #.##

INC

DEC

PGM1 (3)

PGM2 (M)

SET CAL FACTOR

PGM2 (3)

PGM2 (3)

PGM1 (M) PGM2 (M)

PGM1 (3)

BRIDGE VOLTS U/D

INC

DEC

PGM1 (3)

PGM2 (M)

SET BRIDGE VOLTS

PGM2 (3)

CAL FACTOR #.##

PGM1 (M) PGM2 (M)

PGM1 (3)

GAS FACTOR #.##

INC

DEC

PGM1 (3)

PGM2 (M)

SET GAS FACTOR

PGM2 (3)

CAL LEVEL @ ##%

SET CAL LEVEL

Shipping Address: 3200 A-1 Research Forest Dr., The Woodlands, Texas 7381

Mailing Address: P.O. Box 8067, The Woodlands, Texas 77387-8067

phone 888-367-4286, 281-367-4100 • fax 281-292-2860 • www.detcon.com • [email protected]